8.2.1 (HHMI)

Source characteristics
Superbend (5.0 T, single pole)
Energy range
5-17 keV when using standard Si(111) monochromator; 10-13 keV when using multilayer
Monochromator
Double-crystal Si(111) and multilayer
Measured Flux (1.9 GeV, 500 mA)
3.5×1011 ph/s at 12keV through 0.1mm pinhole; 5x flux boost when using multilayer
Resolving power (E/ΔE)
7000
Divergence (max at sample)
3.0 (h) x 0.35 (v) mrad
Spot size
30, 50, or 100 um user-adjustable apertured beam
Endstation
Mini-hutch
Detector
Dectris EIGER2 Si 9M
Automounter Rigaku ACTOR; robot dewar holds 5 ALS-style, Unipucks, or Rigaku ACTOR pucks
Sample format
Hampton or Mitegen ALS pins
Kappa
available; see beamline scientist for use
Special notes
Computers for data processing and analysis available
Full remote operation capability
MD2 installed
Local contact
Marc Allaire
Beamline phone
510-495-2081
def.site file