Pin base warmer is coming along…
BycorieAnthony installed the first pin base warmer on 5.0.2 today. We have to wait until the ALS has LN restored before we can test it. One thing to note: the people we spoke to at NE-CAT about their pin base warmer told us that it does not actually solve the icing issue completely. They still do see pins move on the goniometer after mounting. Hopefully the warmer will help a bit, though!

Pin base warmer looking good
BycorieWe measured the temperature at the sample during warming of the pin base, and preliminary tests so far show that the heat is not transferred to the loop during two minutes of base warming. Tests are continuing now to determine how well the warmer removes ice and frost from the bases.
Images of a crystal in a loop before and after using the pin base warmer:
Shutdown work in progress
BycorieALS is down for the month of July and the first week or so of August. We are taking the opportunity to do a bunch of preventative maintenance on our beamlines. Some of the items include: – cleaning the Be window on 502 (may help to decrease scatter in images) – replacing air lines –…