Source characteristics
|
ALS Sector 5 wiggler (11.4-cm-period) |
Energy Range |
6-17 keV
|
Monochromator |
Double-crystal, Si(111)
|
Measured Flux (1.9 GeV, 500 mA) |
1.5×1012 ph/s at 12keV through 0.1mm pinhole
|
Resolving power (E/ΔE) |
7000
|
Divergence (max at sample) |
2.7 (h) x 0.4 (v) mrad
|
Spot size |
15x15um to 200x200um user-adjustable apertured beam
|
Endstation |
Standard walk-in hutch
|
Detector |
Pilatus3 6M 25 Hz
|
Automounter |
ALS robot; robot dewar holds 12 ALS-style or Unipucks
|
Sample format |
Hampton or Mitegen ALS pins
|
Kappa |
Available; see beamline scientist for use
|
Special notes | Computers for data processing and analysis available Full remote operation capability User-adjustable aperture (CVC) installed |
Local contact | Marc Allaire |
Beamline phone |
510-495-2052
|
def.site file |